Nurse Call Systems • HERO Series

Nurse Call Systems

Medicabay HERO Series wired nurse call systems give patients and staff a reliable communication channel. Call buttons, handsets and WC pull cords in the room connect to a room address module; the room modules share one bus line to the nurse desk panel, with door top lamps in the corridor and Blue Code for life-threatening emergencies.

Key figures
WiredBus system
63Rooms per bus line
9Beds per room
6Call types

Units can be set up to the project before delivery.

Nurse desk

Room and bed number, five calls at once by urgency.

Room module

Works on its own, independent of system faults.

Blue Code

Blue, pink and white codes above every other call.

Lighting

Room and bed head light from the handset.

Hero nurse call desk panel
HERO Series Products

From the bed to the nurse desk

Each component has one job; together they make sure no call is missed.

Hero Nurse Desk Panel
01 / 07 · The central hub at the nurse station

Hero Nurse Desk Panel

The nurse desk panel receives and shows every call: a four-digit room number with the bed number or the call type, and up to five calls at once, ordered by urgency.

Audible and visual alerts change with the call type. The panel also shows the time and room temperature; one master panel can be joined by any number of slave panels, and every system setting is made from the panel.

Rooms
Up to 63 room address modules
Display
Room number, bed number or call type
Calls shown
Five at once, by urgency
Clock & temperature
Built in
Panels
One master, any number of slaves
Code
MGA.NCS.WD.DESK
Hero Room Address Module
02 / 07 · One module per room, on its own

Hero Room Address Module

The room address module connects the units in the room: three bed units, one WC pull cord, one door top lamp and one code call unit, with RJ45 connections.

Microprocessor controlled, it keeps working on its own if the rest of the system has a fault. It mounts in the corridor or in the bed head trunking, is protected against reverse polarity and can be changed while the system runs. Three modules give up to nine beds in one room.

Connects
3 bed units, WC, door lamp, code unit
Beds per room
Up to 9 with three modules
Room connection
RJ45
Bus line
Up to 63 modules
Mounting
Corridor or bed head trunking
Code
MGA.NCS.WD.RMD01
Nurse Call Button & Blue Code
03 / 07 · One press to call for help

Nurse Call Button & Blue Code

The nurse call button at the bed calls the nurse with a single press. The button is lit so it can be found in the dark and glows brighter to confirm the call; the cancel button shows the nurse in the room with a green light.

The emergency button starts Blue, Pink and White Code calls. To prevent false calls it is active only while a nurse is in the room; its red light flashes once the code call has started.

Call button
Lit, confirmation light
Nurse in room
Green light
Codes
Blue, pink and white
Code unit
Active with a nurse in the room
Codes
MGA.NCS.WD.WB01, MGA.NCS.WD.BC01
Hero Handset
04 / 07 · Call and light control from the bed

Hero Handset

The handset plugs into the bed unit and puts the call within reach of the patient.

The red button starts a call; the other two buttons switch the room light and the bed head light through the lighting module.

Call
Red button
Lighting
Room light and bed head light
Connection
Plugs into the bed unit
Code
MGA.NCS.WD.HS01
Hero WC Pull Cord
05 / 07 · Help from the bathroom

Hero WC Pull Cord

Installed in bathrooms, the pull cord lets patients alert staff at once with a red cord that is easy to reach.

Several pull cords can be connected in series to cover every point where a patient may need help.

Cord
Red pull cord
Connection
Several units in series
Call type
Toilet call
Code
MGA.NCS.WD.WB02
Hero Door Top Lamp
06 / 07 · The call status from the corridor

Hero Door Top Lamp

Mounted above the patient door, the door top lamp leads staff to the room that called and shows the call status from a distance.

It signals in red, yellow, green and blue, in an unbreakable housing. Every room has one door top lamp.

Colours
Red, yellow, green and blue
Housing
Unbreakable
Position
Above every patient door
Hero Lighting Control Module
07 / 07 · Room lighting from the handset

Hero Lighting Control Module

The lighting control module switches the room lighting from the handset, so staff and patients need not move around the room, especially during night care.

Microprocessor controlled, it connects to the mains with two relays for two lights.

Control
From the handset
Outputs
Two relays, mains connected
Control
Microprocessor
Code
MGA.NCS.WD.LGH01
Every room: bed units, a room address module, a door top lamp and, where needed, a WC pull cord and a code unit.
Call Types

Six call types, in order of urgency

The nurse desk panel shows the most urgent call first, with its own sound and light.

01

Blue Code

Life-threatening emergencies, started from the code unit; the highest priority.

02

Pink Code

Started from the code unit while a nurse is in the room.

03

White Code

Started from the code unit while a nurse is in the room.

04

Emergency call

Started by the nurse in the room from any call unit.

05

Toilet call

Started with the WC pull cord, with its own sound and light.

06

Normal call

Started with the red button on the bed unit or the handset.

System & Installation

Standards & Technical Specifications for Nurse Call Systems

Common to the HERO Series wired nurse call system.

Network

Wired bus

Room address modules share one bus line to the nurse desk; up to 63 rooms per line.

Power

12 V DC

A central 12 V DC supply; bus on shielded twisted pair, room units on CAT5 / CAT6 with RJ45.

Environment

Up to 50 °C

For hospitals, care homes and similar facilities, up to 50 °C and 90 % humidity.

Quality Management

ISO 13485 & ISO 9001

Supplied under Medicabay’s ISO 13485 medical device and ISO 9001 quality management systems.

HERO Series System Builder

Build your nurse call system

Set the quantities for your ward. The list uses the Medicabay product codes.

1Components and quantities
Hero Nurse Call Desk Panel (MGA.NCS.WD.DESK)Quantity
1
Hero Nurse Call Button (MGA.NCS.WD.WB01)Quantity
8
Hero Nurse Call Hand Set (MGA.NCS.WD.HS01)Quantity
8
Hero Nurse Call WC/Bathroom Pull Cord (MGA.NCS.WD.WB02)Quantity
4
Hero Nurse Call Room Address Module (MGA.NCS.WD.RMD01)Quantity
4
Hero Nurse Call Lighting Control Module (MGA.NCS.WD.LGH01)Quantity
4
Hero Nurse Emergency Button (MGA.NCS.WD.BC01)Quantity
2
Knowledge Base & Specifications

Nurse Call Systems FAQs

Clinical, installation and engineering answers about Medicabay Nurse Call Systems.

What is a nurse call system?

A communication system that lets patients alert nurses or healthcare providers when they need assistance.

Why a wired system?

A wired system avoids the interference and connectivity issues that can occur with wireless systems; each room address module also keeps working on its own.

How many rooms and beds can one system serve?

Up to 63 room address modules on one bus line; with three modules a room can have up to nine beds.

What is Blue Code?

The highest-priority call for life-threatening emergencies, started from the emergency button while a nurse is in the room. Pink and white codes are also available.

Can patients control the room lighting?

Yes. The handset switches the room light and the bed head light through the lighting control module.

Which components are in the system?

Nurse desk panel, call button, handset, WC/bathroom pull cord, room address module, lighting control module and emergency button.

Request a quote

Planning a project with Nurse Call Systems?

Tell us about your facility and our team will prepare a tailored offer and technical documentation.

Scroll to Top