Anaesthetic Gas Scavenging Plant

AGS Series

Medicabay Healthcare Systems introduces Anaesthesia Gas Scavenging (AGS) Plants, engineered to provide a reliable and efficient solution for the safe disposal of anaesthetic gases in healthcare facilities. Designed to help create a safer working environment for medical personnel, our AGS systems effectively remove anaesthetic gas mixtures from operating rooms and areas with nitrous oxide terminal units, reducing exposure to potentially harmful gases and promoting the health and safety of medical staff.

Select Waste Gas Flow Rate
Active Redundant Architecture
70 m³/h 80 m³/h 100 m³/h 150 m³/h 210 m³/h 270 m³/h
Model Code
MGS.AGS.SSF100
Blower Motor
1.1 kW
Max Negative Pressure
-240 mbar
Noise Level
55 dB(A)
Anaesthetic gas scavenging plant (AGSS Series)
Anaesthetic Gas Scavenging Plant

Anaesthetic Gas Scavenging Plant Features & Clinical Safety

Hospital-grade Anaesthetic Gas Scavenging Plant systems engineered to safeguard medical personnel and comply with ISO 7396-2 standards.

Duplex Side-Channel Blowers

Two die-cast aluminium side-channel blowers pre-piped and wired on a common skid. 100% standby redundancy ensures uninterrupted surgical theatre protection.

Automatic Duty Failover

Integrated controls continuously rotate operating hours between blowers. Automatic failover starts the standby pump immediately if differential flow deviates.

Dynamic Vacuum Relief Valve

Factory-calibrated brass relief valve maintains suction pipeline vacuum within safe limits. Prevents excessive negative pressure in surgical breathing circuits.

Spark-Proof Impellers

Side-channel impellers rotate without internal contact or mechanical oil lubrication. Eliminates friction ignition risks during flammable anesthetic vapor transfer.

Remote Theatre Activation

Compatible with BS 6834 / ISO 9170-2 theatre terminal units. Low-voltage remote switches allow surgeons and nurses to activate scavenging on demand.

Pre-Piped Isolation Manifold

Integrated non-return check valves and isolating ball valves ensure each side-channel blower can be serviced without interrupting theatre exhaust.

Airflow & BMS Telemetry

Airflow sensors and differential pressure switches detect blockages or line loss. Dedicated volt-free dry contacts interface seamlessly with hospital BMS.

Pre-Commissioned Compact Skid

Self-contained footprint built on anti-vibration rubber dampers. Shipped pre-tested and ready for immediate drop-in connection to power and piping.

How the AGS Plant works: Operating sequence & P&ID flow

The system is activated directly from theatre remote push buttons, initiating duplex blower suction through dedicated copper pipelines to rooftop exhaust.

OPERATING THEATRE REMOTE BUTTON START / STOP SWITCH 24V REMOTE RUN SIGNAL AGS AGS TERMINAL UNIT BS 6834 / ISO 9170-2 NEGATIVE SUCTION PIPELINE AGS PLANT ROOM SKID CENTRAL CONTROL PANEL DUTY / STANDBY ROTATION DUPLEX SIDE-CHANNEL BLOWERS VACUUM RELIEF VALVE & SENSOR Pressure: -200 to -380 mbar ROOFTOP EXHAUST 0.00 Bar
1 24V Run Signal

Theatre Remote Activation

Surgical staff press the remote push button in the operating theatre, transmitting a low-voltage start signal to the central control panel.

2 -10 to -20 mbar

Terminal Unit Scavenging

Waste gases from the anaesthetic breathing circuit are drawn into the dedicated copper pipeline network through the BS 6834 terminal outlet.

3 -200 to -380 mbar

Duplex Blower Suction

The control panel starts the primary side-channel blower. The vacuum relief valve regulates negative pressure to protect pipeline integrity.

4 0.00 Bar (Atmospheric)

Atmospheric Exhaust

Captured anaesthetic gases are safely discharged through dedicated rooftop exhaust piping, eliminating harmful exposure for hospital staff.

Anaesthetic Gas Scavenging Plant Standards & Safety

Built in strict compliance with international medical gas pipeline standards to protect healthcare professionals from occupational exposure.

Pipeline System

ISO 7396-2 Compliance

Dedicated international benchmark governing medical anaesthetic gas scavenging pipeline systems. Regulates system sizing, minimum terminal extraction flow, and duplex pump redundancy.

UK NHS Standard

HTM 02-01 & BS 6834

UK Department of Health standard defining active theatre scavenging controls, illuminated remote switch operation, and rooftop dispersion safety away from fresh air intakes.

Terminal Units

ISO 9170-2 Gas Outlets

Standard for gas-specific anaesthetic scavenging terminal units. Ensures mechanical non-interchangeability with medical vacuum and guarantees leak-free probe engagement.

Electrical Safety

EN 1012-2 & IEC 60601

Mechanical and electrical safety standard ensuring spark-proof side-channel blower rotation, thermal overload protection, and electromagnetic immunity for central plant room controllers.

AGS Plant Builder ISO 7396-2 • HTM 02-01 • BS 6834

Choose a model by flow

Medicabay AGS Series anaesthetic gas scavenging plants run from 70 to 270 m³/h in 6 standard models. Select your flow capacity and architecture to generate the project tender clause or request an official quote.

Model CodeTypeFlow rate (50 Hz)Max. pressure rangeBlower power (kW)Noise level (dB)Standard

MGS.AGS.SSF100

Waste gas flow
100 m³/h
Architecture
Duplex
Max. pressure range
-240 mbar
Blower power
1.1 kW
Noise level
55 dB(A)
Standard compliance
ISO 7396-2

100% standby redundancy with automated duty changeover and remote operating theatre activation interface.

Frequently Asked Questions

Essential technical, installation and clinical answers regarding Anaesthetic Gas Scavenging Plants.

What is the primary function of an Anaesthetic Gas Scavenging System (AGSS)?

An AGSS actively extracts exhaled anaesthetic gas mixtures from operating theatres. It prevents nitrous oxide and halogenated hydrocarbon exposure to protect surgical and nursing personnel.

Why does the AGS Series utilize duplex side-channel blowers?

Duplex side-channel blowers provide 100% standby redundancy per ISO 7396-2. If the duty blower experiences an issue, the control panel automatically activates the secondary pump without suction interruption.

How is the AGSS activated from surgical operating theatres?

Surgical staff activate the plant via illuminated remote push-button switches installed on theatre walls or pendant arms. The switch transmits a safe 24V low-voltage run signal to the central panel.

Why are side-channel blowers spark-proof and oil-free?

Side-channel impellers rotate without internal metal-to-metal contact or oil lubricants. This non-contact design eliminates friction heat, preventing ignition risks when handling flammable or oxidizing anaesthetic vapor mixtures.

What is the function of the vacuum relief valve in the AGSS pipeline?

The spring-loaded relief valve admits ambient air if line resistance spikes. It maintains pipeline negative pressure within safe limits, preventing excessive suction on patient breathing circuits and anaesthesia ventilators.

How should AGSS rooftop exhaust discharge be positioned?

In accordance with HTM 02-01 and ISO 7396-2, the exhaust must vent above roof level away from building ventilation air intakes, windows, and public access points.